traditional glucometer dc meassurement
@gokul_ragav
1 day ago
https://everycircuit.com/circuit/6314457701679104
trans imepedance opamp circuit :- 0.4 volts from dac is given to non inverting terminal. when we apply blood the resistance decrease and output voltage increases . I strip = I rf = (vout-vx)rf Vx= (r1 / r1+rf)* Vout rf=49.9k r1=100ohms (blood)